TroubleshootingTroubleshooting

Troubleshooting Overview

Where to start when something is wrong with your FizzWizz — triage by symptom, or run the full Health Check if the cause is unclear.

Troubleshooting overview

This section has several different tools. Picking the right one saves time, so start here rather than at the first page in the sidebar.

Start with the symptom

What you are seeingGo to
Carbonation is off target — under-carbonated, or brite pressure behaving oddlyCarbonation & Brite Pressure
You have a failed Health Check ident (A–X)By Symptom (A–X)
An alert fired — CO₂ low, leak, glycol, stone inefficiencyResponding to Alerts
The device is not reporting — Stale or Offline on the dashboardDevice Offline or Not Reporting
Temperature looks wrongSlushy WizzTemp Test
You do not know what is wrongRun the Health Check

When to run the full Health Check

The Health Check is a structured 7-check sequence that isolates problems across the pneumatics, solenoids, sensors, and leaks. It takes about 30 minutes.

Run it when:

  • Commissioning a new or relocated FizzWizz
  • Performing preventative maintenance during off-season or scheduled service
  • After replacing a component (solenoid, transducer, sensor)
  • Carbonation results are off-target with no obvious cause
  • You suspect a leak, stuck solenoid, or faulty sensor
  • Support has asked you for results

A completed Results Form with actual measured values is the single most useful thing you can provide when contacting support. It eliminates most of the back-and-forth.

Read this before replacing parts

Fluid inside the FizzWizz is the most common root cause of transducer and solenoid failures. It happens when the arm separator is not used or is overwhelmed, or when check valves are stuck open or missing.

If several unrelated checks fail, investigate the fluid path before replacing individual components. Replacing a failed transducer or solenoid without establishing how fluid got in will simply reproduce the failure. See Arm Circuit & Separator and the Pneumatics Schematic.

This single principle explains a large share of FizzWizz failures, and it is the reason several pages in this section ask about separator condition and fluid ingress before recommending parts.

The Health Check sequence

CheckWhat it testsIdentsPage
Check 1Setup and power supply—Setup & Power Check
Check 2Display values on CO₂_In circuitA–CCheck 2
Check 3Solenoid function on CO₂_InD–JCheck 3
Check 4Solenoid function on ARMK–QCheck 4
Check 5Display values on ARM circuitR–TCheck 5
Check 6Leak testingU–VCheck 6
Check 7Stone hose and isolation valveW–XCheck 7

Every result is identified by a letter — A through X — that appears consistently across the procedure pages, the Results Form, and the symptom guide.