Health CheckCheck 5 — ARM Display Values

Check 5 — ARM Display Values

Verify CO2 regulator, Brite pressure, and temperature readings with CO2 on the ARM port — Health Check Idents R, S, and T.

Check 5 — ARM display values

With CO2 still connected to the ARM port from Check 4, re-read the same three display values as Check 2. This matters because in Check 2 the arm circuit was open to atmosphere, so Brite psi read zero and a transducer stuck at zero could have passed. Here the arm circuit is pressurized, so Brite psi should now show a real pressure, which is what proves the transducer is actually responding.

Preconditions

  • CO2 still connected to ARM from Check 4
  • Regulator at 30–40 psi
  • FizzWizz in Test Mode

The checks

IdentDisplay valueCriterionTolerance
RCO2 reg psiEquals the supply regulator dial, and falls between 30 and 40 psi±3 psi
SBrite psiEquals the CO2 reg psi reading, and falls between 30 and 40 psi±3 psi
TTemperatureA representative value, similar to the reading recorded at check CJudgement

Reading R: regulator pressure

Same as Ident A. Compare the display's CO2 reg psi against the supply regulator dial, then record both numbers on the Results Form.

Reading S: the key measurement

With gas fed into the ARM port, the arm circuit is pressurized to supply pressure. Brite psi and CO2 reg psi should therefore agree within 3 psi. If Brite psi still reads near zero, the Brite transducer is not responding. Change it and inspect for fluid in the pipe leading to the transducers.

Reading T: compare against C

Temperature should be close to whatever you recorded at check C. A significant change between the two readings, taken minutes apart in the same location, points to an intermittent sensor or a loom fault.

If a check fails

IdentSymptomActions
RWrong reg psi1. Try a different CO2 regulator. 2. Change the CO2_In pressure transducer. 3. Check the CO2_In check valve per Ident L.
SBrite psi reads wrong pressure1. Change the Brite pressure transducer. 2. Inspect for fluid inside the unit.
TTemperature reads wrong1. Change the WizzTemp sensor. 2. Inspect the internal loom. Confirm with the Slushy WizzTemp Test.

Next step

Check 6 — Leak Testing

Test the arm circuit for leaks with the gloved finger and separator tests.